VT1529A/B Verification & Calibration   469
resistor path. Send the MEAS:VOLT:UNStrained? (@<ch_list>) query to 
the VT1422A. The expected return value is ~0.357 V on each channel.
Configure the VT1529A/B for 120
Ω quarter-bridge operation using 
the SENS:FUNC:STR:Q120 (@<ch_list>) command. Send the 
MEAS:VOLT:UNStrained? (@<ch_list>) query to the VT1422A. 
The expected return value is ~0.338 V on each channel.
Configure the VT1529A/B for 350
Ω quarter-bridge operation using 
the SENS:FUNC:STR:Q350 (@<ch_list>) command. Send the 
MEAS:VOLT:UNStrained? (@<ch_list>) query to the VT1422A. 
The expected return value is ~0.301 V on each channel.
Troubleshooting Failures
Check that the Dummy Load is wired and working correctly and firmly 
seated in the RJ-45 input connector. If all 8 channels of one block of 
VT1529A/B channels are failing, check that the Voltage Source is 
connected to the correct terminals on the Bridge Excitation input on the 
VT1529A/B.
Test V-5: Dynamic
Strain Output Port
Description
Test the dynamic output gain and offset of the VT1529A/B.
The dynamic strain output voltage should be equal to V
SENSE
 times the gain 
of the output amplifier, which is fixed at 32.
With V
EXCITATION
 = 5.000 V, the Dummy Load attached and the VT1529A/B 
configured for the 0
Ω sense path, the nominal dynamic output voltage can 
be computed:
for R = 0
Ω,     V
sense
 = 0.357143 V,   V
dynamic
 = 11.4286 V
Setting Vexcitation to 0 V, verify the operation of the dynamic output offset 
control DAC by setting the offset to +1.0 V. The dynamic output voltage 
should be the same as the DAC value, +1.0 V.
Note that the actual measurements will vary based on the actual resistor 
values in the dummy load, so these values should be considered approximate 
when running this test.
Note
This test does not test the offset or bandwidth specifications.
Setup
The VT1529A/B must be connected to the VT1422A through the VT1539A 
SCP. The Dummy Load must be connected to the channel(s) being tested.
The Voltage Source must be connected to the appropriate channel bank 
terminals of the Bridge Excitation input on the VT1529A/B.
The Digital Multimeter should be connected to the first channel on the 
Dynamic Strain Output Port.