SECTION 7—COMPRESSOR
33
Replacing Capacitor
NOTE: For this procedure, refer to FIGURE 7.2 on page 33.
1. Unplug the unit.
2. Remove the cabinet. Refer to Removing Cabinet on page 23.
3. Using needlenose pliers, disconnect the spade connectors from the capacitor
terminals.
4. Cut the tie wrap that secures the capacitor to the base of the concentrator.
5. Remove the capacitor from the concentrator base.
6. Install new capacitor by reversing STEPS 3‐6.
7. Re‐install the cabinet. Refer to Removing Cabinet on page 23.
FIGURE 7.2 Replacing Capacitor
Capacitor
Capacitor Terminals (Spade
Connectors connect here)
Tie Wrap