SECTION 7—COMPRESSOR
34
Rebuilding the Thomas Model 2660 Compressor 
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: This kit is designed for rebuilding the Thomas Model 2660 Series compressor. The model 
number is located on the front of the compressor. Compressor numbers for this kit are 2660CE32, 
2660CE37, and 2660CE50. If the compressor has a different model number than those listed, this 
kit will not work.
The Thomas 2660 Series Compressor Rebuild Kit (p/n 1109235) contains the following: 
DESCRIPTION QUANTITY
Tools Required:
• Torque wrench that has an inch‐pound scale (for head screws, set screws, flapper 
valve screw, and pipe plugs)
• Torx
®
 T‐25 drive for torque wrench (for head screws)
• Torx T‐27 drive for torque wrench (for retainer screws)
• 1/8‐inch Allen wrench attachment for torque wrench (for eccentric set screw)
• 5/32‐inch Allen wrench attachment for torque wrench
• Flat blade screwdriver attachment or 1/4‐inch hex socket for torque wrench (for 
flapper valve screw)
• Two flat blade screwdrivers
• Rubber Mallet
• Phillips screwdriver attachment for torque wrench (for retainer screws)
DESCRIPTION
QUANTITY
Teflon®Piston Cups
2
Aluminum Piston Sleeves
2
Flapper Valves - Intake and Exhaust
4
Valve Keepers
4
O-Ring - Sleeve
2
Valve Plate Gaskets
2
Head Mounting Screws
8
Valve Screws
4
Valve Restraints
2
Piston Cup Retainer Screws
2