SITRANS LR 460 – INSTRUCTION MANUAL 
Page 19
mmmmm
Wi
ri
ng
Wiring
Connecting SITRANS LR 460
1.
Release the cover lock on the enclosure with a 3 mm Allen key and unscrew the 
cover. (Use a screwdriver for extra leverage, if required.)
2.
Loosen the cable gland and push the power cable through until it reaches the 
terminal strip.
 on page 21 for next steps.)
WARNINGS: 
Turn off power to the device before unscrewing the housing cover in a 
hazardous area.
All field wiring for AC models must have insulation suitable for at least 
250 V.
The DC input terminals shall be supplied from a source that provides 
electrical isolation between the input and the output, in order to meet 
the applicable safety requirements of IEC 61010-1.
The equipment shall be protected by a fuse or circuit breaker of up to 
16 A in the building installation.
A circuit breaker or switch in the building installation, marked as the 
disconnect switch, shall be in close proximity to the equipment and 
within easy reach of the operator.
To avoid short-circuits, do not connect a load resistance with bare 
wires inside the connection box.
Notes:
AC and DC input circuits: min. 14 AWG (2.5 sq. mm) copper wire. 
Lay power cables separately from communication wiring.
Recommended torque on terminal clamping screws: 0.5 to 0.6 N m (0.37 to 0.44 Lbf-ft).
power 
cable
cable gland
cover 
lock
metal bracket
terminal 
strip