Wiring
Connecting SITRANS LR 460
1.
Release the cover lock on the enclosure with a 3 mm Allen key and unscrew the
cover. (Use a screwdriver for extra leverage, if required.)
cover. (Use a screwdriver for extra leverage, if required.)
2.
Loosen the cable gland and push the power cable through until it reaches the
terminal strip.
(Go to
on page 20 or
on page 21 for next steps.)
WARNINGS:
•
Turn off power to the device before unscrewing the housing cover in a
hazardous area.
hazardous area.
•
All field wiring for AC models must have insulation suitable for at least
250 V.
250 V.
•
The DC input terminals shall be supplied from a source that provides
electrical isolation between the input and the output, in order to meet
the applicable safety requirements of IEC 61010-1.
electrical isolation between the input and the output, in order to meet
the applicable safety requirements of IEC 61010-1.
•
The equipment shall be protected by a fuse or circuit breaker of up to
16 A in the building installation.
16 A in the building installation.
•
A circuit breaker or switch in the building installation, marked as the
disconnect switch, shall be in close proximity to the equipment and
within easy reach of the operator.
disconnect switch, shall be in close proximity to the equipment and
within easy reach of the operator.
•
To avoid short-circuits, do not connect a load resistance with bare
wires inside the connection box.
wires inside the connection box.
Notes:
•
AC and DC input circuits: min. 14 AWG (2.5 sq. mm) copper wire.
•
Lay power cables separately from communication wiring.
•
Recommended torque on terminal clamping screws: 0.5 to 0.6 N m (0.37 to 0.44 Lbf-ft).