SITRANS LC 500 – INSTRUCTION MANUAL
Page 29
m
mmmm
Int
erc
onnect
ion
Note: 
Grounding the SITRANS LC 500 as illustrated above provides only system 
grounding for referencing purposes: it does not provide safety grounding. 
Cathodically 
Protected Metal 
Tanks
Cathodically protected metal tanks 
are never directly grounded. 
However, the impedance of the 
supply source is so low that it will 
not cause any problems. The 
shielding of the loop cable should 
be grounded at one end only (the 
tank end) to avoid short-circuiting 
the cathode protection voltage.
The ground lug on the 
SITRANS LC 500 can be connected 
to the tank as shown. 
(See page 27 for further grounding 
details.)
metal
ground lug
optional stilling 
well
V
KP
Non-Conductive Tanks
Non-metallic tanks always require a 
stilling well or proper grounded 
conductive medium.
Connect the ground lug on the 
SITRANS LC 500  to  earth  ground.  If 
the stilling well is integral to the 
SITRANS LC 500, it is now grounded.
If the vessel has a stilling well 
provided, make sure that the metal 
parts of the stilling well are properly 
grounded.
ground lug
stilling well
non-
metallic