SITRANS LC 500 – INSTRUCTION MANUAL
Page 35
m
mmmm
Int
erc
onnect
ion
• The Upper Sensor Limit (USL) and Lower Sensor Limit (LSL) are set to 3300 and 
1.666 pF respectively, and the following conditions apply: the Upper Range Value 
(URV) and Lower Range Value (LRV) should be within the USL to LSL range, but 
can be set anywhere within that range.
• An interruption of the measuring connection will be detected: a loose or 
interrupted connection results in up to 0.5 pF capacitance, which is below the 
adjusted LSL and thus signals a FAULT condition.
DATE
"customer input data via HART"
SENSOR SERIAL NUMBER
"customer input data via HART"
FINAL ASSEMBLY NUMBER
"customer input data via HART"
TV1   Units 
UNDEFINED
TV1   LRV 
0    [switch position (0)E, TV1]
TV! URV
1.0 [switch position (0)F, TV1]