Omega OMB-DAQ-2416_4AO 사용자 설명서

다운로드
페이지 53
OMB-DAQ-2416-4AO User's Guide 
Specifications 
27  
Connect thermocouples to the OMB-DAQ-2416-4AO such that they are floating with respect to GND (pins 18, 
36, 39, 42, 44, 46, 66, 81, 84, 93, and 94).  
When configuring thermocouple sensors, keep any stray capacitance relative to GND (pins 18, 36, 39, 42, 44, 
46, 66, 81, 84) as small as possible to avoid settling time and accuracy errors. 
The OMB-DAQ-2416-4AO GND and DGND pins are isolated from earth ground. You can connect 
thermocouple sensors to voltages referenced to earth ground as long as isolation between the GND/DGND pins 
and earth ground is maintained.  
Table 4. J type thermocouple accuracy specifications, including CJC measurement error. 
All specifications are (±).  
A/D data rate 
Sensor temperature range 
Accuracy error  
maximum 
Accuracy error  
typical 
Tempco  
(°C/°C) 
3750 S/s 
 
-210 °C 
2.572 °C 
1.416 °C 
0.022 
0 °C  
0.935 °C 
0.469 °C 
1200 °C 
1.869 °C 
1.456 °C 
2000 S/s 
 
-210 °C 
2.572 °C 
1.416 °C 
0.022
 
0 °C  
0.935 °C 
0.469 °C 
1200 °C 
1.869 °C 
1.456 °C 
1000 S/s 
-210 °C 
2.572 °C 
1.416 °C 
0.022
 
0 °C  
0.935 °C 
0.469 °C 
1200 °C 
1.869 °C 
1.456 °C 
500 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
100 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
60 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
50 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
25 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
10 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
5 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
2.5 S/s 
-210 °C 
2.442 °C 
1.334 °C 
0.022
 
0 °C  
0.881 °C 
0.415 °C 
1200 °C 
1.821 °C 
1.408 °C 
To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for 
45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416-
4AO is housed in its plastic enclosure.