Table 5. K type thermocouple accuracy specifications, including CJC measurement error. 
All specifications are (±). 
A/D data 
rate 
Sensor temperature 
range 
Accuracy error 
maximum  
Accuracy error 
typical 
Tempco 
(°C/°C) 
3750 S/s 
-210 °C 
2.917 °C 
1.699 °C 
0.029
 
0 °C 
1.017 °C 
0.526 °C 
1372 °C  
2.478 °C 
2.022 °C 
2000 S/s 
-210 °C 
2.917 °C 
1.699 °C 
0.029 
0 °C 
1.017 °C 
0.526 °C 
1372 °C  
2.478 °C 
2.022 °C 
1000 S/s 
-210 °C 
2.917 °C 
1.699 °C 
0.029 
0 °C 
1.017 °C 
0.526 °C 
1372 °C  
2.478 °C 
2.022 °C 
500 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
100 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
60 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
50 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
25 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
10 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
5 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
2.5 S/s 
-210 °C 
2.735 °C 
1.524 °C 
0.029 
0 °C 
0.948 °C 
0.457 °C 
1372 °C  
2.396 °C 
1.941 °C 
To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for 
45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416-