Invacare Compressor 사용자 설명서

다운로드
페이지 40
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS
11
If any leakage of the oxygen cylinder is detected DO NOT 
attempt to use the cylinder. Turn the flow selector knob on the 
regulator to Off. If leakage persists, place the cylinder 
outdoors and notify your Home Health Care Provider or 
Service Representative of this condition.
The HomeFill II is equipped with a high pressure relief valve 
to ensure the user’s safety. When activated, this safety feature 
will make an extremely loud noise. If this noise occurs, turn 
the unit Off and contact your Home Health Care Provider or 
Service Representative.
Unplug unit from outlet when not in use and before cleaning.
This appliance is provided with double insulation. Use only 
identical replacement parts. Refer to Servicing of 
Servicing of Double-Insulated Products
In a double-insulated product, two systems of insulation are 
provided instead of grounding. No grounding means is 
provided on a double-insulated product, nor is a means for 
grounding to be added to the product. Servicing a 
double-insulated product requires extreme care and 
knowledge of the system, and is to be done only by qualified 
service personnel. Replacement parts for a double-insulated 
product MUST be identical to the parts they replace. A 
double-insulated product is marked with the words 
"DOUBLE INSULATION" or "DOUBLE INSULATED". The 
symbol ( ) is also able to be marked on the product. 
CAUTION
DO NOT unplug the power cord by pulling on the cord. To 
unplug, grasp the plug not the cord, otherwise damage may 
occur.
SAVE THESE INSTRUCTIONS