SECTION 5—INDICATOR LIGHTS
32
SECTION 5—INDICATOR LIGHTS
INDICATOR LIGHT 
COLOR
COMPRESSOR 
STATUS
EFFECT
None
Compressor is Off
O
2
 flowing to user if 
concentrator is on
O
BELOW NORMAL - 
YELLOW
Compressor is On and 
warming up 
(0-3 minutes after 
power is On)
Cylinder filling not 
started
O
2
 flowing to user if 
concentrator is On
O
2
 from concentrator 
is < 90% (any time after 
three minute warmup)
Cylinder filling stops, 
will restart when O
2
 
> 90% 
O
2
 flowing to user if 
concentrator is On
FILLING - GREEN
Compressor is On and 
cylinder is filling. O
2
 
from concentrator is 
>90% (any time after 
three minute warmup)
Cylinder is filling
O
flowing to user if 
concentrator is On
FULL - GREEN
Compressor is On. 
Cylinder is finished 
filling (full) -remove 
cylinder (FULL light is 
On in >10 mins.) 
or 
Cylinder is not 
attached properly - 
check connection 
(FULL light is On in 
<10 mins.)
Cylinder filling stops
Cannot fill another
cylinder until 
compressor is turned 
Off and On again
O
2
 flowing to user if 
concentrator is On