SECTION 5—INDICATOR LIGHTS
33
ATTENTION - RED 
(Audible Alarm)
Compressor is On. 
Cylinder NOT filling
Causes:
Full cylinder attached 
when unit turned On - 
replace with empty 
cylinder
or
Internal failure - 
contact dealer
Compressor does not 
begin to fill or stops 
filling 
Unit must be turned Off 
and On before 
filling can occur
O
2
 flowing to user if 
concentrator is On
INDICATOR LIGHT 
COLOR
COMPRESSOR 
STATUS
EFFECT