Gossen R2600 사용자 설명서

다운로드
페이지 40
GOSSEN METRAWATT GMBH
17
3.4 Request for event data
The event date, combined in 2 words, contain all error messages and alarms of the equipment. 
They can be called up via short set to identify a special error. 
This request can be made asynchronous, for example, if the  service request bit (collected errors) was set before in any random response telegram in 
the function field (FF).
Example: Equipment address = 5
See 2.4.2.2 for contents of the function field (FF)
The 4 characters of the event data block are bit fields which are combined to form the error status words 1 and 2. 
These 4 characters can also be read by data request with parameter index 21h. 
See the operating instructions for more explanations and information on error elimination.
Request from master (short set):
10h
05h
A9h
AEh
16h
R2600/01 response (long set, see 2.3):
68h
06h
06h
68h
05h
„FF“
Data block
„PS“
16h
4 characters