A&D AD-4212B-101/102/201/301 사용자 설명서
82
23. SPECIFICATIONS
AD-4212A-100 AD-4212A-200 AD-4212A-600
AD-4212A-1000
Weighing capacity
110 g
210 g
610 g
1100 g
Maximum display
110.0084 g
210.084 g
610.084 g
1100.084 g
Minimum weighing value (1 digit)
0.1 mg
1 mg
Repeatability (Standard deviation)
0.15 mg
1 mg
Linearity
±
0.3 mg
±
2 mg
±
3 mg
0-30 g
1.1
0-30 g
0.8
0-30 g
0.9
0-30 g
0.9
Stabilization time in seconds
(typical at FAST under good
environment)
30-110 g 1.3 30-210 g 1.0 30-610 g 1.1
30-1100 g
1.1
Display refresh rate
10 times/second (5 times/second can be selected)
RS-232C
Bi-directional, 600-19200 bps
Comparator contact
output
3-level output
I/O unit
External contact input
RE-ZERO, PRINT
Buzzer
Sounds the connected buzzer for 3-level comparison results
(It becomes 5-level when OP-04 is installed)
Sensitivity drift
±2 ppm/°C (10°C-30°C/50
°F-86°F )
Operating environment
5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation)
Calibration weight provided
(Conforming to OIML Class E2)
50 g
100 g
200 g
Applicable weight values
20 g, 50 g, 100 g
20 g, 50 g,
100 g, 200 g
20 g, 50 g, 100 g,
200 g, 300 g, 400 g,
500 g, 600 g
20 g, 50 g, 100 g, 200 g,
300 g, 400 g, 500 g, 600 g,
700 g, 800 g, 900 g, 1000 g
Dimensions 80
(W)
×
230 (D) × 90 (H) mm
Weighing pan
50 × 50 mm
70 × 70 mm
Weighing
unit
Mass Approx.
2
kg
Dimensions 237
(W)
×
150 (D) × 155 (H) mm
Display
Mass Approx.
2
kg
Connection cable
Approx. 2 m (between weighing unit and display)
AC adapter
Confirm that the adapter type is correct for the local voltage and power receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter )
Minimum unit mass
0.1 mg
1 mg
Counting
mode
Number of samples
10, 25, 50 or 100 pieces
Minimum 100%
reference mass
10.0 mg
100 mg
Percent
mode
Minimum 100%
display
0.01%、0.1%、1%
(Depends on the reference mass stored.)
OP-01
BCD output, external contact input
(
RS-232C and comparator contact output disabled)
OP-04 RS-232C,
5-level
comparator
contact output, external contact input
(Not field installable. Must be ordered with the balance for installation at the factory.)
OP-07* Extension
cable
3
m
OP-08 Ethernet
interface
(
RS-232C and comparator contact output disabled)
OP-19
Stainless steel breeze break (Provided as standard for AD-4212A-100)
Options
OP-20 Metal
leveling
foot
Only one extension cable can be used. The maximum distance between the weighing unit
and the display unit is 5 m (connection cable 2 m + extension cable 3 m).
and the display unit is 5 m (connection cable 2 m + extension cable 3 m).