4.  Zero Adjustments 
 
It is impossible for there to be any zero drift so zero adjustments are 
never required.  
 
5.  Recalibration 
 
Please contact the Omega Engineering Service Department if your flow 
sensor or flow meter needs recalibration. 
 
6.  Using Flow Sensors or Flow Meters with Different Gases 
 
The FLR 1000 Series will operate with most gases subject to compatibility 
of the wetted parts. 
 For information about calibrating for Hydrogen or 
Helium, contact the Omega Engineering Service Department.
 
 
Units are calibrated with AIR at a pressure of 760mmHg and temperature 
of 23°C unless otherwise detailed on the calibration certificate. Using other 
gases will effect the calibration. The amount of calibration error will 
depend on the characteristics of the gas being flowed. 
 
The flow reading correction factor (or calibration factor) may be 
approximately calculated with the following formula: 
 
 
 
C
f
 = ( S
new
 ÷ S
current
 )
0.3
 
 
C
f
 = Correction factor 
S
new
 = Specific gravity of new gas 
S
current
 = Specific gravity of original calibration gas 
 
 
 
A corrected flow reading may then be calculated using the correction 
factor as follows: 
 
 
Corrected Flow = Flow reading X Correction Factor 
 
 
For example, if the unit is calibrated for AIR but the user wants to 
measure the flow of Carbon Dioxide (CO
2
) with a specific gravity of 1.517 
the flow reading correction factor would be calculated as follows: 
 
 
S
new
 = Specific gravity of CO
2
 = 1.517 
S
current
 = Specific gravity of Air = 1.000 
 
C
f
 = (S
current
 ÷ S
new
)
0.3
 
 
C
f
 = ( 1.000 ÷ 1.517 )
0.3
 
 
C
f
 = 0.88