D.
 
Maintenance and Product Care 
 
1.  General  
 
 
CAUTION: Do not disassemble your unit for any reason. 
If the unit appears to be malfunctioning please contact 
the Omega Engineering Service Department. 
 
 
The FLR 1000 Series of flow meters and sensors require no periodic 
maintenance if used within the recommended specifications. 
 
The next recommended recalibration date is stated on the calibration 
certificate supplied with the unit. Please contact the Omega Engineering 
Service Department for further information. 
 
Inlet filters should be periodically checked and cleaned / replaced as 
necessary. 
 
Regularly check all electrical and process connections for damage or 
deterioration. 
 
If the sensor is to be stored, keep both the inlet and outlet ports sealed.  
 
 
2.  Cleaning and Flushing 
 
If there is a build up of deposits or particles from the measured gases it 
may be necessary to clean 
or flush
 the unit. This should be done by 
flowing clean, particle free air (or Nitrogen) through the unit at a flow rate, 
pressure and temperature within the specifications of the unit. If 
necessary, flow may be reversed to assist flushing. 
 
 
CAUTION: Do not flow a high pressure gas through the 
sensor. A burst of high pressure gas may cause the 
turbine wheel sensor to be damaged. Always flow gas at 
a pressure and flow rate within the specifications of the 
unit 
 
 
3.  Returning Units for Repair or Recalibration 
 
To return a unit for repair or recalibration please contact the Omega 
Engineering Customer Service Department. An Authorized Return (AR) 
number will then be issued. The AR number should then be noted on the 
outside of the package and on any correspondence. Further details may be 
found on page 28 of this manual.