Chapter 4 Ventilator Setup
61
Connect Supplemental Oxygen (Optional)
To attach supplemental oxygen to the device:
1.  Connect the oxygen tubing to the Oxygen Inlet Quick Connect. 
Then, connect the Quick Connect to the Oxygen Inlet on the 
back of the ventilator.
Oxygen Inlet
Oxygen Source
Tubing
Oxygen Inlet
Quick Connect
2.  Make sure you press the Oxygen Inlet Quick Connect firmly into 
the inlet. You should hear a click, and the latch at the top of the 
inlet will pop up.
 
This device is only compatible with a low flow oxygen source 
providing up to 15 l/min.  
Attaching the O
2
 Tubing
Press Tubing in and Latch
Pops Up 
 CAUTION
Do not connect an unregulated 
or high pressure oxygen source 
to this connector on the device.