Trilogy200 clinical manual
62
Connect the Remote Alarm (Optional)
You can use a remote alarm as a nurse call system or in-house remote 
alarm system. It can generate an alarm at a distance of up to  
91 meters (300 feet) from the ventilator. An audible tone sounds 
and a red light blinks to indicate that an alarm condition exists. The 
remote alarm will sound when any of the following conditions exist:
 
The ventilator is off.
 
Any alarm occurs and is not silenced or reset.
Refer to the Accessories chapter of this manual for information on 
which remote alarm systems are compatible with the device. Refer to 
the instructions included with the Remote Alarm Adapter Cable for 
information on how to connect a remote alarm to the ventilator.
 WARNING
When using a remote alarm, 
make sure you fully test the 
remote alarm connector and 
cable by verifying that:
-   Annunciated  alarms on 
the ventilator are also 
annunciated on the remote 
alarm.
-   Disconnecting the remote 
alarm cable from the 
ventilator or from the remote 
alarm results in an alarm 
notification at the remote 
alarm.
 WARNING
The remote alarm should be 
tested daily.