Shutting your appliance down
If your appliance is to be shut down for any length of time, switch 
it off and disconnect the plug or switch off or unscrew the fuse. 
Clean the appliance and leave the door open in order to prevent 
unpleasant smells. 
Malfunctions
You may be able to rectify the following faults by checking 
the possible causes yourself:
•  Appliance does not function:
– Is the appliance switched on?
– Is the plug correctly fitted in the mains socket?
– Is the fuse intact?
•  Loud running noise:
– Is the appliance set up firmly on the floor?
– Does the appliance cause nearby items of furniture or objects to 
vibrate? Please note that noises caused by the refrigerant circuit 
cannot be avoided.
•  The temperature is not low enough:
– Is the temperature setting correct (see "Setting the temperature")?
–  Does the separately installed thermometer show the correct read-
ing?
–  Is the ventilation system working properly?
–  Is the appliance set up too close to a heat source?
• 
 appears in the display:
– Reset the real time clock (see "Setting the real time clock").
If  none  of  the  above  causes  ap-
ply  and  you  cannot  rectify  the 
fault yourself, contact the nearest 
customer  service  department 
stating  the  type  designation 
1
service  number 
2
and  appliance 
number 
3
  as  indicated  on  the 
type plate. 
Terminating resistor
When connecting several appli-
ances via the RS485 interface, the 
terminating resistor must remain 
on the last appliance.
Remove the terminating resis-
tors from the appliances in 
between.
External alarm
We recommend connecting the appliance 
to an external alarm device.
There are various connection options at 
the back of the appliance.
The appliance may only be connected 
to an external alarm device by trained 
personnel.
Floating alarm output
These three contacts can be used to connect the appliance to an 
optical or acoustic alarm device.
The connection is designed for 250 V / 8 A AC or 36 V / 8 A DC.
Floating alarm 
output
Connection to 
RS485 interface
Note
The  connectors  are  secured  with 
screws. To remove the connectors, 
undo the left and right screws.
COM
External power supply unit
250 V AC (alternating current) or
36 V DC (direct current - positive pole)
N.C
Operating light
Connection for a control lamp to indicate 
that the appliance is in normal mode.
N.O 
Alarm output
Connection for a visual warning light or 
an acoustic alarm signal.
Optional temperature 
sensor connection