3M 245 사용자 설명서

다운로드
페이지 23
English
15
202524A
3M
TM 
Ranger
TM
 Blood/Fluid and Irrigation Warming Systems
Check the 3M
TM 
Ranger
TM
 system website to ensure you have the most recent version of this document. 
www.rangerfluidwarming.com reorder #202524A
Check the 3M
TM 
Ranger
TM
 system website to ensure you have the most recent version of this document. 
www.rangerfluidwarming.com reorder #202524A
Ranger Blood/Fluid Warming Unit, Model 245 Maintenance Log
The operating temperature and alerts should be tested at least every 6 months or according to 
institutional protocol. 
 
m
AintenAnCe
 A
Ction
 p
eRfoRmed
 
model
  ___________ 
Sn
 ________
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
Date  _________
 
Operating temperature:
 (Specification = 41°C±1) 
______°C
 
Over-temperature alert points:
  
 
Audible 
(43°C [±1] on Ranger unit display) 
______°C
 
Secondary 
(46°C [+3, -2])
 ______°C
M
od
el
 2
45