Ranger Irrigation Fluid Warming System, Model 247 Preventative 
Maintenance Checklist
c
aution
Temperature adjustments cannot be made to the Ranger irrigation fluid warming unit. If the test 
shows the temperature is out of specification, please call 3M Patient Warming technical support at 
1-800-733-7775.
t
oolS
 
and
 
equipment
•  Thermocouple: 0.005" lead or smaller  
(part #90020)
•  Calibrated thermocouple temperature monitor
•  Aluminum tape
•  Ranger irrigation fluid warming set
•  Phillips screwdriver
•  Saline/water
1.  Use aluminum tape to affix a thermocouple 
to the top surface of the Ranger irrigation 
fluid warming cassette as shown in Fig. 1. 
2.  Slide the warming cassette into the slot of 
the Ranger irrigation fluid warming unit.
3.  Using room temperature saline/
water, prime the cassette per 
instructions included with it.
note
To achieve correct results, 
fluids must be administered into the 
inlet tube, NOT into the outlet tube.
4.  Close patient line with white clamp 
to stop flow.
5.  Turn the Ranger irrigation 
warming unit ON and allow it to 
warm up for 5 minutes. 
6.  The Ranger irrigation warming 
unit display should read 41°C±1. 
Note the temperature on the 
thermocouple temperature 
monitor. If the temperature reading 
is not 41°C±1, call 3M Patient 
Warming technical support.
note
: Most handheld temperature  
meters have a ±1°C tolerance.
Fig. 1: Ranger irrigation fluid warming cassette
0.75"
8.50"
Thermocouple
Inlet tube
Aluminum tape
 I
rri
ga
tio
n F
lu
id W
ar
m
in
g
D
ra
in 2
m
L o
r m
or
fro
m c
as
se
tte t
o r
em
ov
e.
In
ser
t ca
sse
tte
 
be
fo
re
 p
rimin
g.
M
od
el 24
7
Fig. 2: Ranger irrigation fluid warming set
Exit
White 
clamp
Blue clamp
Fluid source
Cassette
Inlet tube
Outlet tube
M
od
el
 2
47