 Blood/Fluid and Irrigation Warming Systems
English
22
Check the 3M
TM 
Ranger
TM
 system website to ensure you have the most recent version of this document. 
www.rangerfluidwarming.com reorder #202524A
Specifications
phySiCAl
 
ChARACteRiStiCS
Warming
 
unit
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long; wt.: 7 lb. 7 oz. (3.4 kg)
certiFicationS
 IEC/EN 60601-1; CAN/CSA-C22.2, No. 601.1
claSSiFication
MEDICAL — GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND 
MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, 
No. 601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No. 4HZ8
 Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, 
Type B, Ordinary equipment, Continuous operation. Classified by Underwriters Laboratories Inc. with 
respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN
2
 60601-1 and in 
accordance with Canadian/CSA C22.2, No. 601.1. Classified under the Medical Device Directive as a 
Class IIb device.
leAkAge
 
CURRent
Meets leakage current requirements in accordance with UL/IEC 60601-1. 
enviRonmentAl
 
ConditionS
operating
 
temperature
 
range
15° to 40°C (59° to 104°F) 
Storage
 
temperature
 
range
-20° to 45°C (-4° to 113°F)
operating
 
humidity
10 to 85% RH, noncondensing
atmoSpheric
 
preSSure
 
range
50 kPa to 106 kPa
eleCtRiCAl
 
ChARACteRiStiCS
tempeRAtURe
 
ChARACteRiStiCS
device
 
rating
100-120 VAC, 50/60 Hz 
220-240 VAC, 50/60 Hz
Set
 
point
 
temperature
41°C
maximum
 
heating
 
poWer
900 W
over
-
temperature
 
alert
Model 245:
43°C
Model 247:
48°C
FuSe
 
2 x T10A-H (250V) for 100-120 VAC 
2 x T6.3A-H (250V) for 220-240 VAC
under
-
temperature
 
alert
33°C
FuSe
 
type
 
Time delay, high breaking
over
-
temperature
 
cutoFF
Model 245:
44°C (warming units Rev N and newer) 
46°C (warming units Rev A to M)
Model 247: 
50°C