Troubleshooting
CONDITION
CAUSE
SOLUTION
Nothing illuminates on the 
warming unit panel.
• Unit is not turned on, plugged
in, or power cord is not plugged 
into an appropriate outlet.
• Unit failure.
• Turn unit on. Make sure the power
cord is plugged into the power 
entry module of the warming unit. 
Make sure the warming unit is 
plugged into a properly grounded 
outlet.
• Check fuses.
• Call 3M Patient Warming technical
support.
CONDITION
CAUSE
SOLUTION
Alert indicator illuminates and 
alert sounds, alphanumeric 
display alternately flashes a 
temperature of 43˚C for the Model 
245 (48°C for the Model 247) or 
higher and the word “HI.”
Temporary over-temperature 
condition because:
• An extreme change in flow rates
occurred (e.g., from 500 mL/min 
to stop flow).
• Unit was turned on and reached
set point temperature before 
cassette was inserted.
• Fluids were prewarmed to above
42°C before being run through 
the warming unit.
• Open flow to reduce temperature. 
Alerts will stop when the display 
reads 41°C. The unit is ready to use.
• Alerts will stop when the display
reads 41°C. The unit is ready to use.
• Turn off unit. Discontinue infusion
of fluids. Do not warm fluids before 
infusing them through the Ranger 
warming unit.
CONDITION
CAUSE
SOLUTION
Alert sounds, alphanumeric 
display and alert indicator light 
go dark.
Primary controller failure. Unit 
will no longer operate.
Power to heating plates will shut 
off if temperature rises to 44°C for 
the Model 245 (50°C for the Model 
247). Turn unit off and unplug it. 
Discontinue use of unit. Discard 
warming set. Alert will continue 
to sound if you do not unplug unit. 
Call 3M Patient Warming technical 
support.