To clean Wave Plus system components use a soft cloth damped with neutral soap. System components 
are NOT protected by liquid infiltration.  
 
The cleaning of the system elements has to be done: 
1.  Base unit: every six months or more often if required  
2.  Electrodes: every 50 hour or more often if required 
 
Warning:  
 
Liquid  infiltration  can  cause  the  EMG  signal  amplitude  reduction,  or  the  interruption  of  the  electrode 
functioning. In these cases, proceed as follow: 
 
Take  off  the  acquisition  module  (upper  part)  from  the  battery  module,  as  described  in 
“Replacement of electrode battery” chapter; 
Remove the liquid with a dry cloth;  
Dry out the modules using a low temperature source (radiator or oven at maximum 30°C); 
 
If the problem persists, replace the electrode.  
 