To clean Wave Plus system components use a soft cloth damped with neutral soap. System components
are NOT protected by liquid infiltration.
are NOT protected by liquid infiltration.
The cleaning of the system elements has to be done:
1. Base unit: every six months or more often if required
2. Electrodes: every 50 hour or more often if required
Warning:
Liquid infiltration can cause the EMG signal amplitude reduction, or the interruption of the electrode
functioning. In these cases, proceed as follow:
functioning. In these cases, proceed as follow:
-
Take off the acquisition module (upper part) from the battery module, as described in
“Replacement of electrode battery” chapter;
-
Remove the liquid with a dry cloth;
-
Dry out the modules using a low temperature source (radiator or oven at maximum 30°C);
If the problem persists, replace the electrode.