AngioDynamics UK Limited AS0670 사용자 설명서

다운로드
페이지 78
DELTA 15/30 
DELTA 15/30 Operator Manual 
Page 3.3 
Version B.2 
 
 
TO ENSURE BEST PERFORMANCE AND EFFICACY, ALWAYS USE 
DIOMED FIBERS, PROCEDURE KITS AND ACCESSORIES. 
  
ONLY 600
µµµµ
m NON-CONTACT FIBERS WITH A COOLING SHEATH CAN BE 
CALIBRATED. 
 
CONTACT FIBERS DO NOT REQUIRE CALIBRATING. 
 
THE DISPLAYED POWER WITH A CONTACT FIBER OR UNCALIBRATED 
NON-CONTACT FIBER IS THE POWER LEVEL AT THE LASER APERTURE. 
IT SHOULD BE ASSUMED THAT THE POWER LEVEL AT TISSUE IS 10-15% 
LOWER. 
 
 
11.  The system will automatically go to the STANDBY mode, with a set of 
default operating parameters. If required, adjust these parameters now 
using the SCROLL/CONFIRM control. The DIOMED DELTA 15/30 is 
now ready to begin the treatment. 
 
12.  To start treatment and delivery of laser energy, press 
STANDBY/READY, wait for the DIOMED DELTA 15/30 to enter READY 
mode and depress the footswitch.  During laser radiation the laser 
emission indicator will light and an audible warning will be heard. 
 
13.  To pause treatment, release the footswitch. To continue treatment, press 
the footswitch. To end treatment, release the footswitch and return the 
unit to STANDBY. 
 
 
14.  A summary of laser energy delivered may be reviewed if required by 
selecting Statistics from the Main Menu
 
15.  To turn the DIOMED DELTA 15/30 OFF turn the key switch and remove 
the key, then switch off the rear power switch. 
 
If an error message is displayed, refer to section 5 - Technical Information.