Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Changing Parameter and System Setup Settings
The initial settings on the PARAMETER SETUP and SYSTEM SETUP screens can only be 
changed before monitoring. Changing these settings during monitoring interrupts monitoring.
NOTE
Changing Parameter and System Setup settings must be done by qualified personnel.
Changing PARAMETER SETUP Settings
Parameter Setup Setting List
The factory default settings are underlined.
Setting item
Description
Settings
ECG ELECTRODES Select the electrode lead type.
IEC, AHA
LEAD TYPE
Select the type of ECG leads.
AUTO, 6 LEADS
ECG 
MEASUREMENT
Turn ECG monitoring on or off. When electrodes 
are attached to the patient and ECG leads are 
connected, ECG monitoring starts even when this 
setting is set to OFF.
If this setting is set to OFF, the same setting on the 
receiving monitor must also be set to OFF.
NOTE
When “PROTOCOL” on the transmitter is set 
to 51 and the receiving monitor is able to 
receive protocol 51, ECG measurement on 
the receiving monitor is automatically set to 
OFF when this setting is set to OFF on the 
transmitter. 
To use protocol 51, the ORG-9100A/ORG-
9110A multiple patient receiver software 
version 03-03 or later is required.
ON, OFF
RESP 
MEASUREMENT
Turn respiration monitoring on or off.
When this setting is set to OFF, the same setting on 
the receiving monitor is automatically set to OFF.
ON, OFF