Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Changing Parameter Setup Settings
ECG ELECTRODES
Select the electrode lead type.
1.  On the PARAMETER SETUP screen, press 
the 
▼ key to move the cursor to “ECG 
ELECTRODES”.
2.  Press the Function key. The cursor moves to 
the selection item.
3. Press 
the 
▼ key to select “IEC” or “AHA”.
4.  Press the Function key to register the 
selected setting. The cursor returns to “ECG 
ELECTRODES”.
LEAD TYPE
Select the type of ECG leads. In normal use, select “AUTO”. When using DIN type lead with 6 
electrodes, select “6 LEADS”.
1.  On the PARAMETER SETUP screen, press the 
▼ key to move the cursor to “LEAD TYPE”.
2.  Press the Function key.
3. Press 
the 
▼ key to select “AUTO” or “6 LEADS”.
4.  Press the Function key to register the selected setting. The cursor returns to “LEAD TYPE”.