ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
When not monitoring SpO
2
, disconnect the SpO
2
 
cable from the transmitter. Otherwise, noise from 
the probe sensor may interfere and incorrect data 
is displayed on the screen.
Turn off the power of mobile phones, small 
wireless devices and other devices which produce 
strong electromagnetic interference around a 
patient (except for devices allowed by the hospital 
administrator). Radio waves from devices such as 
mobile phones or small wireless devices may be 
mistaken as pulse waves or respiration waves and 
the displayed data may be incorrect.
CAUTION
Handle the probe cable according to the following 
cautions. Failure to follow these cautions may 
cause cable discontinuity or short circuit of the 
probe cable which may cause incorrect 
measurement data or inability to perform 
measurement. Also in rare cases, the probe 
temperature may increase and cause skin burn on 
the patient. If the probe cable is damaged, replace 
the probe with a new one.
‡ 'RQRWSXOORUEHQGWKHSUREHFDEOH
‡ 'RQRWOHWFDVWHUIHHWUXQRYHUWKHSUREHFDEOH
CAUTION
Normal external light does not affect monitoring but 
strong light such as a surgical light or sunlight may 
affect monitoring. If affected, cover the measuring 
site with a blanket.