ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
Close the battery case cover during operation. If 
the transmitter is used with the battery case cover 
open, anyone who touches the opened battery 
case may receive an electrical shock when 
defibrillation is performed. Touching the opened 
battery case may cause electrostatic discharge 
and intermittently interfere with the waveform or 
data.
WARNING
When the signal is unstable, keep the patient 
under close observation. When the signal is 
unstable, the monitoring and alarm are not reliable 
and the receiving monitor cannot detect a sudden 
change of the patient’s condition. This may cause 
critical changes in the patient condition to be 
overlooked. Install an appropriate antenna system 
to ensure stable signal condition.
WARNING
While the “ALARMS SUSPENDED” message is 
displayed on the transmitter, all alarms on the 
receiving monitor are suspended so keep the 
patient under close observation.
WARNING
When the patient returns to the bed, turn on the 
transmitter and check that the monitoring is 
resumed on the receiving monitor.
WARNING
If the transmitter is not turned off and monitoring 
continues for the selected interval, pause 
monitoring is canceled and monitoring continues. 
Check that the monitoring is resumed on the 
receiving monitor.