ZM-540PA, ZM-541PA Operator’s Manual
WARNING
Close the battery case cover during operation. If
the transmitter is used with the battery case cover
open, anyone who touches the opened battery
case may receive an electrical shock when
defibrillation is performed. Touching the opened
battery case may cause electrostatic discharge
and intermittently interfere with the waveform or
data.
the transmitter is used with the battery case cover
open, anyone who touches the opened battery
case may receive an electrical shock when
defibrillation is performed. Touching the opened
battery case may cause electrostatic discharge
and intermittently interfere with the waveform or
data.
WARNING
When the signal is unstable, keep the patient
under close observation. When the signal is
unstable, the monitoring and alarm are not reliable
and the receiving monitor cannot detect a sudden
change of the patient’s condition. This may cause
critical changes in the patient condition to be
overlooked. Install an appropriate antenna system
to ensure stable signal condition.
under close observation. When the signal is
unstable, the monitoring and alarm are not reliable
and the receiving monitor cannot detect a sudden
change of the patient’s condition. This may cause
critical changes in the patient condition to be
overlooked. Install an appropriate antenna system
to ensure stable signal condition.
WARNING
While the “ALARMS SUSPENDED” message is
displayed on the transmitter, all alarms on the
receiving monitor are suspended so keep the
patient under close observation.
displayed on the transmitter, all alarms on the
receiving monitor are suspended so keep the
patient under close observation.
WARNING
When the patient returns to the bed, turn on the
transmitter and check that the monitoring is
resumed on the receiving monitor.
transmitter and check that the monitoring is
resumed on the receiving monitor.
WARNING
If the transmitter is not turned off and monitoring
continues for the selected interval, pause
monitoring is canceled and monitoring continues.
Check that the monitoring is resumed on the
receiving monitor.
continues for the selected interval, pause
monitoring is canceled and monitoring continues.
Check that the monitoring is resumed on the
receiving monitor.