ZM-540PA, ZM-541PA Operator’s Manual
When monitoring respiration is needed, measure
respiration with an instrument. The transmitter
calculates SpO
respiration with an instrument. The transmitter
calculates SpO
2
of arterial blood based on the
principle of pulse oximeter and does not measure
respiration.
respiration.
CAUTION
• Do not use the same channel for different
patients. If the same channel is used for two
patients, the two patients’ data will be lost due to
mutual modulation interference, or another
patient’s data may appear on the receiving
monitor screen.
patients, the two patients’ data will be lost due to
mutual modulation interference, or another
patient’s data may appear on the receiving
monitor screen.
• Do not use two transmitters with adjacent
channels in the same hospital. If transmitters
with adjacent channels are used, their radio
waves interfere with each other.
with adjacent channels are used, their radio
waves interfere with each other.
CAUTION
When monitoring SpO
2
only, detection of
arrhythmia and asystole is not available and
arrhythmia alarms such as ASYSTOLE, VF or VT
are not available. If the patient requires ECG
monitoring, monitor the ECG.
arrhythmia alarms such as ASYSTOLE, VF or VT
are not available. If the patient requires ECG
monitoring, monitor the ECG.
CAUTION
When monitoring SpO
2
only (without ECG
monitoring), turn on both the upper and lower limit
alarms for PR and SpO
alarms for PR and SpO
2
on the receiving monitor. If
the patient’s pulse is not detected during asystole
or other condition, a “CANNOT DETECT PULSE”
or “CHECK PROBE” alarm occurs instead of an
SpO
or other condition, a “CANNOT DETECT PULSE”
or “CHECK PROBE” alarm occurs instead of an
SpO
2
limit alarm. Furthermore, if the patient has no
pulse, noise from probe movement could be
misjudged as a pulse and cause an incorrect PR or
SpO
misjudged as a pulse and cause an incorrect PR or
SpO
2
value to be displayed.