ZM-540PA, ZM-541PA  Operator’s Manual
Always lock the battery cover while using the 
transmitter. Otherwise the battery cover may come 
off and correct measurement might not be 
performed.
CAUTION
Output Signal
WARNING
Do not use the output signal from the receiving 
monitor as the synchronization signal for other 
equipment such as IABP, MRI, echocardiography 
or defibrillator. There may be time delay between 
the monitor and the other equipment caused by 
waveform transmission delay and spike noise may 
interfere on the output signal and be mistaken as a 
trigger.
Battery
Battery replacement must be performed by the 
operator. When replacing the batteries of a 
transmitter that is currently used for a patient, 
disconnect the electrode leads from the transmitter 
before replacing batteries or do not touch the 
patient during replacement.
CAUTION
Refer to the battery and battery charger manuals 
for details on handling the batteries.
CAUTION