ZM-540PA, ZM-541PA  Operator’s Manual
Attaching the Probe to the Patient
$WWDFKWKHSUREHWRWKHSDWLHQWE\UHIHUULQJWRWKHSUREH¶VPDQXDO0DNHVXUHWKDWWKHOLJKW
HPLWWHUDQGSKRWRGHWHFWRURIWKHSUREHIDFHHDFKRWKHUDWWKHDWWDFKPHQWVLWH
WARNING
•  When using the TL-201T finger probe, do not 
fasten the probe and cable to the finger by 
wrapping with tape. This may cause burn, 
congestion or pressure necrosis from poor blood 
circulation.
•  When using probes other than the TL-201T 
finger probe, to avoid poor circulation, do not 
wrap the tape too tight. Check the blood 
circulation condition by observing the skin color 
and congestion at the skin peripheral to the 
probe attachment site. Even for short-term 
monitoring, there may be burn or pressure 
necrosis from poor blood circulation, especially 
on neonates or low birth weight infants whose 
skin is delicate. Accurate measurement cannot 
be performed on a site with poor peripheral 
circulation.
WARNING
Check the circulation condition by observing the 
skin color at the measurement site and pulse 
waveform. Change the measurement site every 8 
hours for disposable probes and every 4 hours for 
reusable probes (every 8 hours for TL-631T series 
probe). The skin temperature may increase at the 
attached site by 2 or 3°C (4 or 5°F) and cause a 
burn or pressure necrosis. When using the probe 
on the following patients, take extreme care and 
change the measurement site more frequently 
according to symptoms and degree.
•  Patient with a fever
•  Patient with insufficient peripheral circulation
•  Neonate or low birth weight infant with delicate 
skin