ZM-540PA, ZM-541PA Operator’s Manual
When Using YP-810P series Disposable Cuffs or YP-702T/703T or YP-960T series
Reusable Cuffs
Reusable Cuffs
To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required.
NOTE: Connect the joints properly. If there is an air leak, NIBP cannot be measured
properly.
1
Connect the NIBP cuff to the extension hose.
2
Connect the other end of the extension hose to
the NIBP socket on the transmitter. Turn the joint
clockwise until it clicks.
the NIBP socket on the transmitter. Turn the joint
clockwise until it clicks.
To disconnect the cuff from the transmitter, turn the
hose joint counterclockwise.
hose joint counterclockwise.
Attaching the NIBP Cuff to the Patient
WARNING
Be careful when measuring NIBP on a patient with
known bleeding disorders or coagulation. After
NIBP measurement, there may be dot
hemorrhage, or circulatory disorder by thrombus
where the cuff is attached.
known bleeding disorders or coagulation. After
NIBP measurement, there may be dot
hemorrhage, or circulatory disorder by thrombus
where the cuff is attached.
Do not wrap the cuff on an arm or thigh which is
used for injection. NIBP measurement on an arm
or thigh which is used for injection may cause
reflux of blood and stop injection.
used for injection. NIBP measurement on an arm
or thigh which is used for injection may cause
reflux of blood and stop injection.
CAUTION