ZM-540PA, ZM-541PA  Operator’s Manual
When Using YP-810P series Disposable Cuffs or YP-702T/703T or YP-960T series 
Reusable Cuffs
To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required.
NOTE:  Connect the joints properly. If there is an air leak, NIBP cannot be measured 
properly.
1
 
Connect the NIBP cuff to the extension hose.
2
 
Connect the other end of the extension hose to 
the NIBP socket on the transmitter. Turn the joint 
clockwise until it clicks.
To disconnect the cuff from the transmitter, turn the 
hose joint counterclockwise.
Attaching the NIBP Cuff to the Patient
WARNING
Be careful when measuring NIBP on a patient with 
known bleeding disorders or coagulation. After 
NIBP measurement, there may be dot 
hemorrhage, or circulatory disorder by thrombus 
where the cuff is attached.
Do not wrap the cuff on an arm or thigh which is 
used for injection. NIBP measurement on an arm 
or thigh which is used for injection may cause 
reflux of blood and stop injection.
CAUTION