ZM-540PA, ZM-541PA  Operator’s Manual
Do not attach the cuff to the site where there is 
injury or inflammation. If the skin gets irritated or 
redness appears on the skin from the cuff, change 
the attachment site or stop using the cuff. Take 
extreme care on the patients with delicate skin.
CAUTION
Do not wrap the cuff too tight. It may cause poor 
blood circulation and congestion. If the cuff is 
wrapped too loosely, the NIBP value may increase.
CAUTION
Do not reuse disposable parts and accessories.
CAUTION
NIBP and SpO
2
 can be measured on the same 
limb, but the SpO
2
 monitoring might not be 
accurate during NIBP measurement. Be careful 
when reading the SpO
2
 values.*
CAUTION
 0RQLWRULQJ6S2
2
GXULQJ1,%30HDVXUHPHQW 
:KHQWKH6S2
2
SUREHLVDWWDFKHGWRWKHVDPHOLPEDVWKH1,%3FXIIWKHEORRGÀRZ
GHFUHDVHVGXULQJ1,%3PHDVXUHPHQWDQGSXOVHZDYHFDQQRWEHGHWHFWHGDQG6S2
2
 cannot 
EHPRQLWRUHGSURSHUO\:KHQ³,1+,%,76S2
2
'85,1*1,%3´RQWKH3$5$0(7(5
6(783VFUHHQLVVHWWR21IDFWRU\GHIDXOWVHWWLQJ6S2
2
PRQLWRULQJLVSDXVHGGXULQJ1,%3
PHDVXUHPHQWWRDYRLG6S2
2
DODUPRFFXUUHQFH+RZHYHUZKHQPRQLWRULQJ6S2
2
 on the same 
OLPEDVWKH1,%3EHFDUHIXOZKHQUHDGLQJ6S2
2
YDOXHV