  The XpandOrtho XO
1
 knee balancing system is contraindicated in partial knee replacements such 
as unicompartmental arthroplasty.  
  The XpandOrtho XO
1
 knee balancing system may not be used in cases with severe deformity: for 
example; a residual varus or valgus greater than 5° even after making bone cuts and ligament 
releases.   
  The XpandOrtho XO
1
 knee balancing system may not be used in cases with severe mismatch in 
the gap at flexion and that at extension (flexion-extension gap mismatch greater than 5 mm).   
  The XpandOrtho XO
1
 knee balancing system is contraindicated in cases in which the minimum 
gap between tibiofemoral surfaces is less than 8 mm.   
  The XpandOrtho XO
1
 knee balancing system may not be used when the controller cannot be 
safely anchored to the thigh: for example, when the distance between the superior edge of the 
incision and the inferior margin of the tourniquet is too small to place the controller securely.   
Precautions: 
  Read and follow Instructions for Use (IFU) for the correct application of the XO
Device, GUI and 
displayed data. 
  Data displayed is for reference purposes only and should be used at the surgeo ’s dis retio  a d 
should not be the sole basis for surgical decisions. 
  Check that the gap between the femur and tibia is at least 8mm.   
  Insert the tibiofemoral distractor in a fully deflated condition only.   
  Do not forcibly insert the tibiofemoral distractor between the femur and tibia.  Do not forcibly 
strike any part of the XO
1
 Device or the attached tibiofemoral distractor with any instrument 
(such as a mallet).   
  The XO
1
 Device is intended to be used for approximately 15 minutes during surgery.  The device 
will not operate properly once the Controller Status Light flashes red.   
 
 
    
If the Controller Status Light flashes red stop using the device immediately.    
  The XO
Device is intended to be used by surgeons; skilled, trained and knowledgeable in the 
performance of total knee arthroplasty (TKA). 
  The XO
Device should be used in a sterile operating room environment only. 
  Use only the provided computer, software, CD ROM, and wireless receiver with the XO
Device. 
  Ensure the PC is in Airplane Mode and the Wi-Fi is turned off 
 
Warnings 
 
  WARNING: No modification of this equipment is allowed. 
  Do not attempt to inflate the XO
1
 device with any means other than those supplied with the 
XpandOrtho XO
1
 knee balancing system. 
  Do not use The XO
1
 device with components other than those supplied in the XpandOrtho 
package. 
  Do not attempt to inflate the XO
1
 device with any liquid. 
  Do not attempt to re-process or re-sterilize the XO
1
 device. 
  The pouch containing the tray is not a sterile barrier.