Varex Imaging Deutschland AG XRPAD4336 사용자 설명서

다운로드
페이지 35
 
w w w . p e r k i n e l m e r . c o m  
X R p a d ™   4 3 3 6   M E D                   2 6  
 
 
 
 
 
 
 
 
 
 
 
 
U
S
E
R
 
M
A
N
U
A
L
 
9
 
Inspection and Maintenance 
  
WARNING
 
The XRpad™ 4336 MED detector must be repaired by PerkinElmer 
authorized personnel only. Ignoring this warning may result in explosion, 
fire, electric shock, or unknown hazards, which may result in severe personal 
injury, death, or substantial product damage. 
  
Caution
 
Inspect the XRpad™ 4336 MED detector before use. In addition, carry out 
prescribed, regular inspections per the instructions in this manual. 
It is important that the detector is used safely and as intended. Inspect the detector and its accessories 
before use. If any problem is found during the inspection, correct the problem, and take measurements 
indicated in this chapter. If the problem cannot be corrected, contact your dealer, distributor, or any 
PerkinElmer subsidiaries (regional service headquarters) listed on the last page of this document  
We recommend that you keep records of the inspection close to the detector. You can make copies of 
the checklist in this chapter or make your own checklist. 
9.1
 
Daily Inspection  
Perform the following inspection daily. If there is any problem, inform your establishment safety 
representative immediately to contact your dealer, distributor, or PerkinElmer subsidiary. 
9.1.1
 
Before Turning ON the Power 
 
Inspection 
Result 
Remedy 
Date 
Date 
Date 
ab
le
Check all cables (Power and 
communication tethered cord, DC-
cable, Ethernet cable, Sync cable) to 
ensure that they are not damaged and 
the insulation is not damaged. 
Good/Bad 
Good/Bad 
Good/Bad 
Contact your dealer, distributor, or 
any PerkinElmer subsidiaries if there 
is a problem. 
Check all connector plugs and locks to 
ensure they are not loose. 
Good/Bad 
Good/Bad 
Good/Bad 
Fully insert the cables and lock them. 
D
et
ec
to
Check that the detector is not 
damaged. 
Good/Bad 
Good/Bad 
Good/Bad 
Contact your dealer, distributor, or 
any PerkinElmer subsidiaries if there 
is a problem. 
Check that the Battery Pack is not 
damaged. 
Good/Bad 
Good/Bad 
Good/Bad 
Replace the Battery Pack with a new 
one. 
Check that the detector is not loose and 
all screws are fixed. 
Good/Bad 
Good/Bad 
Good/Bad 
Contact your dealer, distributor, or 
any PerkinElmer subsidiaries if there 
is a problem. 
9.1.2
 
After Turning ON the Power 
 
Inspection 
Result 
Remedy 
Date 
Date 
Date 
G
en
er
al
 
Check that the wireless connectivity 
symbol (  ) is shown in the display if 
the Wireless Mode is used. 
Good Bad 
Good/Bad 
Good/Bad 
Connect the Detector and the WiFi 
Access Pointer  as described in the 
Access Pointer Manual 
Check that the LAN connectivity 
symbol (        
) is shown in the display if 
the Wired Mode is used. 
Good Bad 
Good/Bad 
Good/Bad 
Connect the Gigabit Ethernet cable 
and the tethered power and 
communication cable properly.