Huntleigh Healthcare Limited Huntleigh Diagnostic Products Division SF1US 사용자 설명서

다운로드
페이지 48
 11
 Pr
oduct Inf
or
ma
tion
3.2.2 
Front Panel Indicators
Docking Transducers 
B
C
A
A - Docking/Charge Bays
Ensure there is no water or gel on the contact panels of the receiver or 
transducer when charging, as this may prevent good contact. 
Dock the transducers in position A when not in use. Place the transducers with  
the golden charging rings facing the charging pins and snap into place. 
The transducers are equipped with re-chargeable lithium polymer batteries. 
When the transducer is docked the battery is charged automatically, provided 
the receiver is connected to the mains supply. 
Once a transducer has been positioned correctly for charging, a Green LED in 
the corresponding holder (B) turns on.
It doesn’t matter in which of the two positions a transducer is docked. However, 
to ensure the transducers are recognised without delay, the Ultrasound 
transducer should be placed in the holder fi rst. If the TOCO transducer is 
inserted fi rst, there may be a delay in the transducer being recognised.
Note:  
The gold plated contact rings MUST be clean & dry before the 
   
transducer is docked.  Failure to ensure this may cause  
 
   
corrosion, poor contact and may invalidate the warranty.