Olympus Winter & Ibe GmbH ESG200 사용자 설명서
101
Appendix B
APPENDIX B
Mode characteristics according to IEC 60601-2-2
Mode characteristics of monopolar modes
The waveforms of all monopolar modes are continuous, except of ForcedCoag
1)
.
Mode
Default
power
setting
Minimum
power
level
3)
Maximum
power
level
3)
Maximum
output
voltage
Rated
load
End-of-
procedure
load
FineCut
30
5 W
120 W
580 V
p
500 Ω
—
PureCut
30
5 W
120 W
630 V
p
500 Ω
—
StrongCut
30
5 W
120 W
680 V
p
500 Ω
—
Pulse Cut
30
5 W
120 W
690 V
p
500 Ω
—
SoftCoag
30
5 W
120 W
250 V
p
100 Ω
—
ForcedCoag
2)
30
5 W
120 W
2000 V
p
500 Ω
—
Mode characteristics for monopolar modes
1)
Waveform ........................................................................... Modulated sinus, 15 kHz
2)
Crest factor at rated load ............................................................................3.0 to 8.0
3)
The power level is the continuous power (Watts) which can be applied during the
activation of the electrosurgical generator. The actual applied power depends on the
tissue characteristics, e.g. resistance.
Mode characteristics of bipolar modes
The waveforms of all bipolar modes are continuous.
Mode
Default
power
setting
Minimum
power
level
1)
Maximum
power
level
1)
Maximum
output
voltage
Rated
load
End-of-
procedure
load
FineCut
30
5 W
120 W
470 V
p
500 Ω
—
PureCut
30
5 W
120 W
530 V
p
500 Ω
—
StrongCut
30
5 W
120 W
560 V
p
500 Ω
—
SoftCoag
30
5 W
120 W
260 V
p
100 Ω
—
Fine RFITT
10
1 W
40 W
250 V
p
100 Ω
dynamic
Pure RFITT
10
1 W
40 W
150 V
p
100 Ω
400 Ω
Strong RFITT 20
1 W
40 W
260 V
p
100 Ω
900 Ω
DRAFT
–
2017-01-17