Beha Amprobe GT-800 STD KITVDE-tester 4472062 사용자 설명서
제품 코드
4472062
GT-600/GT-800 Instruction Manual
72
Frequency range (20 mA)
30 ... 723 Hz (characteristics according to DIN
EN 61010-1 Annex A, fig. A.1 and VDE 0411
Part
1)
Clamp type
AC, input max. 60 A, output 1mA/A
(Type CHB1, No. 2390055)
Clamp Current TRMS (
) / Range 0...60 A (GT-800 only)
Display range
0.2 ... 60.0 A
Resolution 0.1
A
Accuracy (without clamp error) +/- (5% rdg. + 3 digits)
Input resistance
Input resistance
0.5
Frequency range
30 ... 723 Hz
Clamp type
AC, input max. 60 A, output 1mA/A
(Type CHB1, No. 2390055)
IEC Cord Test (CORD)
Measurements / test steps
- R
PE
protective bonding test (used R
PE-5A
function, limit value adjustable 0.01 ... 10.00
), accuracy of R
PE-5A
measurement
(5% rdg.
+ 6 digits)
-
-
Riso insulation resistance test L/N against PE
insulation 0.00 ... 20.0 M
(used R
ISO-500V
function, fixed limit value 1.00 M
), accuracy
of R
ISO-500V
measurement
10% (0.25 ... 20
M
).
-
L/N shorted - pass/fail (used R
ISO-250V
function,
limit value 1 M
).
-
N continuity - pass/fail (used R
ISO-250V
function, fixed limit value approx. 1 M
).
-
L continuity - pass/fail (used R
ISO-250V
function,
fixed limit value approx. 1 M
).
Cord R
PE
test lead compensation Yes, 0.00 ... 5.00
Pre-Tests and Protection,
Functions and Messages
Functions and Messages
Refer to page 64
PRCD Test (GT-800 only)
Type of PRCD
Type of PRCD
AC (2-pole, 3-pole and type S)
Nominal differential current I
N
Selectable 10 mA or 30 mA
Measurements / Test steps -
Visual
test
-
R
PE
(used R
PE-5A
function, limit value
adjustable 0.01 ... 10.00
), accuracy of R
PE-5A
measurement
(5% rdg. + 6 digits)
-
R
ISO
output side (used R
ISO
500V
function, fixed
limit value 1 M
), accuracy of R
ISO-500V
measurement
10% (0.25 ... 20 M).
-
I
PE
(used I
PE(
)
function, fixed limit value 3.50
mA, accuracy +/- (5% rdg. + 3 digits))
-
Functional test (test button, PRCD must trip)