5-5
CH 5/ Start Menu
Automatic Setup 
procedure 
(continued)
Action...
Result...
STEP 5: Select your desired 
probe type.
NOTE: Use the same probes for 
the following: TR2500 for 
TR2500A, TR2510 for TR2510A, 
TR2520 for TR2520A. 
• If the current probe that you are 
using is not on the list, select 
Other
. Then press Scale to 
define the CT scale factor on 
the current probe. Proceed to 
Step 6.
• Press OK to accept probe 
selection and monitor current.
• Press Cancel to retain previous 
probe settings and return to 
previous screen.
EP104a
STEP 6: You need to set the CT 
scale factor when using a current 
probe that is not on the list. 
Pressing Scale will show the 
numeric window to enter the CT 
scale factor on the current probe. 
Use the keypad to set the CT scale 
factor. See sample scale factor 
calculation on page 5-10.
• Press OK to accept the CT 
scale factor value. The Current 
Probe Selection screen will be 
displayed.
• Press 
...
 
....
 if you want to clear 
and retype numeric values.
• Press x1000 to multiply the 
value by 1000.
• Press Cancel to discard 
changes and return to the 
previous screen (screen shown 
on Step 5).
EP104d