Advanced bqMTester Instructions
6. Select which stations will have their internal or external temperature probes available for use for
calibration by clicking the On Board Temp Probe or Ext Temp Probe checkbox. If neither internal nor
external temperatures are selected, then that station will be required to either use a temperature probe
from another station or manual input of the temperature.
NOTE:
These selections are not to be confused with the temperature sensors that are part of the
fuel gauge module being tested. These are part of the HPA495 test/calibration equipment.
7. Select the Use for Test checkbox to enable a station for use during testing. If the Use for Test is
de-selected, then that station will be disabled and will not perform testing. A disabled station
s
temperature probe will be available to other stations; however, if it is selected from Step 5 in this
section.
8. Clicking Flash LED for each station causes the corresponding calibration board to flash its LEDs and
enable the current and voltage power supplies. This is useful for testing the power supplies and for
identifying the corresponding hardware for each station.
9. Click Save and Proceed.
Figure 5. Multi-Station Setup
4.5
Temperature Probe Selection
Clicking Save and Proceed brings up the Temperature Probe Selection window. This window is used to
configure the temperature probes. For each station, you can select either of the following options:
1. "No Tracking
Use entered value."
2. Any on-board or external temperature probe that is selected at station setup stage (the Use for Test
checkbox must be checked for that station, see
will be available in the drop-down box menu
list.
For example:
16
Advanced bqMTester
SLUU397A
February 2011