Advanced bqMTester Instructions
Figure 6. Global Configuration Screen
4.6.1
Board Offset Calibration
Board Offset Calibration is activated depending on which device is selected. Some devices do not
support Board Offset Calibration for each module tested. If activated, this option is used to check for board
offset values within user configurable limits. The software automatically computes the board offset in
microvolts for the cases where it is activated and enabled. You can change these values, if needed.
4.6.2
CC Offset Calibration
CC Offset Calibration is the coulomb counter offset. There are no user-definable values in this box. This
calibration can be selected by checking the box or deselected by unchecking the box. The default is that
the box is checked.
NOTE:
If this test is disabled, the values from the Golden Image File will be used and not the
values currently in the part.
4.6.3
Voltage Calibration
Voltage Calibration can be selected by placing a check in its selection box, or deselected by removing
the check. The default is checked. The voltage calibration area also has a box to enter the number of
series cells being simulated. The default number of cells is four. If this checkbox is not selected, the
corresponding voltage frame is grayed out.