70
Repair procedures
Welch Allyn Propaq CS Vital Signs Monitor
P1 removal
6.
Spread the casing open to an angle of about 45-60 degrees.
7.
Detach the following cables from the Front Chassis (shown below):
Cable to P5 on the Analog Board, from left side panel ECG connector on the Rear 
Chassis.
Cable to P6 on the Analog Board, from left side panel IPB connector on the Rear 
Chassis.
Cable to P7 on the Analog Board, from left side panel temperature jack on the 
Rear Chassis.
Cable to P8 on the lower (Digital) Board, from the right side panel on the Rear 
Chassis.
Cable to P9 on the Digital Board, from the right side panel on the Rear Chassis.
P6
P7
P5
P11
P10
P9
P7
P5
P6
P8
NOTE: You do not need to remove 
the handle for this procedure. The 
handle is not shown here so the 
cable connections are more visible