VT1422A Command Reference   273
CALibration
CALibration:TARE?
CALibration:TARE? 
Returns a value to indicate the success of the last CAL:TARE 
operation. CAL:TARE? returns the value only after the CAL:TARE operation 
is complete.
Returned Value: 
The data type for this returned value is int16.
Executing CAL:TARE sets the Calibrating bit (bit 0) in Operation Status Group. 
Executing CAL:TARE? resets the bit.
Related Commands: CAL:STOR TARE
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Command
Sequence
CAL:TARE  <ch_list>
to correct channel offsets
CAL:TARE?
to return the success flag from the 
CAL:TARE operation
CAL:STORE  TARE
Optional depending on necessity of long 
term storage
CALibration:VALue:RESistance 
CALibration:VALue:RESistance  <ref_ohms
sends the just-measured value of 
the on-board reference resistor to the module for A/D calibration.
Parameters
Comments
Use the CAL:CONF:RES command to configure the reference resistor for 
measurement at the Calibration Bus connector.
A four-wire measurement of the resistor can be made with an external multimeter 
connected to the HCALLCALHOHM and LOHM terminals on the Terminal 
Module or the V HV L
Ω H, and Ω L terminals on the Cal Bus connector 
when not using a terminal module.
Value
Meaning
Further Action
0
Cal OK
None
-1
Cal Error
Query the Error Queue (SYST:ERR?)
Also run *TST?
-2
No results available
Perform CAL:TARE
before CAL:TARE?
Parameter
Name
Parameter
Type
Range of
Value
Default 
Units
ref_ohms
numeric (float32)
7,500 ± 5%
ohms