VT1422A Command Reference   275
CALibration
<ref_volts> may be specified in millivolts (mV).
 When Accepted: Not while INITiated
Related Commands: CAL:CONF:VOLT, CAL:STORE ADC
Send with VXIplug&play Function: hpe1422_cmd(...)
Command
Sequence
CAL:CONF:VOLTAGE  4,FSCALE
*OPC?
Wait for operation to complete
enter statement
(now measure voltage with external DMM)
CAL:VAL:VOLT  <measured value>
Send measured value to module
CALibration:ZERO?
CALibration:ZERO? 
corrects Analog to Digital converter offset for any drift since 
the last *CAL? or CAL:ZERO? command was executed. The offset calibration takes 
about 5 seconds and should be done as often as the control set up allows.
Comments
The CAL:ZERO? command only corrects for A/D offset drift (zero). Use the 
*CAL? common command to perform on-line calibration of channels as well as 
A/D offset. *CAL? performs gain and offset correction of the A/D and each 
channel with an analog SCP installed (both input and output).
Returned Value: 
The data type for this returned value is int16.
Executing this command does not alter the module's programmed state 
(function, range, etc.).
Related Commands: *CAL?
*RST Condition: A/D offset performed
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage
CAL:ZERO?
enter statement here
returns 0 or -1
Value
Meaning
Further Action
0
Cal OK
None
-1
Cal Error
Query the Error Queue (SYST:ERR?)