VXi VT1422A 사용자 설명서

다운로드
페이지 529
VT1422A Command Reference   337
[SENSe]
[SENSe:]DATA:FIFO:PART?
[SENSe:]DATA:FIFO:PART?  <n_values
returns <n_values> from the FIFO 
buffer.
Parameters
Comments
Use the DATA:FIFO:COUNT? command to determine the number of values in 
the FIFO buffer.
The format of values returned is set using the FORMat[:DATA] command.
Returned Value: ASCII values are returned in the form 
±1.234567E±123. 
For example 13.325 volts would be +1.3325000E+001. Each value is followed 
by a comma (,). A line feed (LF) and End-Or-Identify (EOI) follow the last value. 
The data type is a string array.
REAL 32, REAL 64 and PACK 64, values are returned in the IEEE-488.2-1987 
Definite Length Arbitrary Block Data format. This data return format is 
explained in “Arbitrary Block Program and Response Data” on page 233. 
For REAL 32, each value is 4 bytes in length (the data type is a float32 array). 
For REAL 64 and PACK 64, each value is 8 bytes in length (the data type is a 
float64 array).
Note
Values which are a positive overvoltage return IEEE +INF and a negative overvoltage 
return IEEE -INF (see Table 7-1 on page 294 for actual values for each data format).
Related Commands: DATA:FIFO:COUNT?
*RST Condition: FIFO buffer empty
Use VXIplug&play function: hpe1422_readFifoFast_Q(...)
Usage
DATA:FIFO:PART?  256
return 256 values from FIFO
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
n_values
numeric (int32)
1 - 2,147,483,647
none