VT1422A Command Reference   343
[SENSe]
[SENSe:]FUNCtion:CUSTom:TCouple
[SENSe:]FUNCtion:CUSTom:TCouple <type>,[<range>,](@<ch_list>) 
links channels with the custom Engineering Unit Conversion table loaded with the 
DIAG:CUST:PIECE command. The table is assumed to be for a thermocouple and 
the <type> parameter will specify the built-in compensation voltage table to be used 
for reference junction temperature compensation. SENS:FUNC:CUST:TC allows an 
EU table to be used that is custom matched to thermocouple wire that has been 
characterized. Contact a VXI Technology System Engineer for more information on 
Custom Piecewise Engineering Unit Conversion for specific applications.
Parameters
Comments
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4 V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 generates an error. Specifying 
0 selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range.
If using amplifier SCPs, set them first and keep their settings in mind when 
specifying a range setting. For instance, if the expected signal voltage is to be 
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8, 
<range> must be set no lower than 1 V dc or an input out-of-range condition 
will exist.
The <type> EEXTended applies to E type thermocouples at 800
°C and above.
The *CAL? command calibrates temperature channels based on Sense Amplifier 
SCP setup at the time of execution. If SCP settings are changed, those channels 
are no longer calibrated. *CAL? must be executed again.
Related Commands: DIAG:CUST:PIEC, *CAL?,SENS:REF, and 
SENS:REF:TEMP
*RST Condition: all custom EU tables erased
Send with VXIplug&play Function: hpe1422_cmd(...)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
type
discrete (string)
 E | EEXT | J | K | N | R | S | T
none
range
 numeric (float32)
see comments
V dc
ch_list
channel list (string)
100 - 163
none