VXi VT1422A 사용자 설명서

다운로드
페이지 529
344 VT1422A Command Reference  
[SENSe]
Usage
program must put table constants into array table_block
DIAG:CUST:PIEC 1,table_block,(@100:107)
send characterized thermocouple table 
for use by channels 0-7
SENS:FUNC:CUST:TC N,.25,(@100:107)
link custom thermocouple EU with chs 
0-7, use reference temperature 
compensation for N type wire.
SENSE:REF  RTD,92,(@120)
designate a channel to measure the 
reference junction temperature
include these channels in a scan list (REF channel first)
INITiate then TRIGger module
[SENSe:]FUNCtion:FREQuency
[SENSe:]FUNCtion:FREQuency (@<ch_list>) 
sets the SENSe function to 
frequency for channels in <ch_list>. Also configures the channels specified as digital 
inputs.
Parameters
Comments
If the channels specified are on an SCP that doesn't support this function, 
an error will be generated. See the SCP's User's Manual for its capabilities.
Use the SENSe:FREQuency:APERture command to set the gate time for the 
frequency measurement.
Related commands: SENS:FREQ:APER
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital 
SCP channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:FUNC:FREQ (@144)
set channel 44's to frequency
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ch_list
string
100 - 163
none